Suggested Immediate Loading Protocol for AIDI Dental Implants

----Dr. William Y.S. Hung 4/17/2010

Warning: This protocol only applies to 90% of cases, because every case has to be evaluated according to occlusion, missing teeth and medical condition. You are encouraged to contact AIDI for evaluation before you load AIDI Dental Implants.

  1. You may take immediate impression after you place AIDI Dental Implants if the initial stability exceeds 45 Ncm.
  2. Generally, AIDI only suggests immediate non-functional load for both anterior & post restorations with proper canine guidance & good posterior support.
    1. Single unit anterior restorations - initial stability must between 80 -90 Ncm.
    2. Single unit molar restorations (2 implants required) - initial stability must be 75 - 80 Ncm.
    3. 3 units - initial stability exceeds 70 Ncm of all implants for all anterior teeth implant restoration and/or if initial stability exceeds 80 Ncm for all posterior teeth implants.
    4. 4 units of anterior teeth restoration - initial stability exceeds 70 Ncm for 3 implants and the remaining one dental implants exceeds 50 Ncm.
    5. 4 units of posterior teeth restoration - initial stability exceeds 80 Ncm for 3 implants and initial stability of the remaining one dental implants exceeds 50 Ncm.
    6. Cross arch splinting for mandibular arch - initial stability of 4 dental implants exceed 70 Ncm, the remaining implants exceed 50 Ncm.
    7. Cross arch splinting for maxillary arch - initial stability of 5~6 dental implants exceed 70 Ncm, the remaining implants exceed 50 Ncm.
  3. The above protocol only applies to AIDI dental implant system.
  4. Please note, DO NOT TORQUE EXCEED 90Ncm